Fuji Electric ZSVS-2 Instruction Manual page 93

Compact type gas analyzer
Table of Contents

Advertisement

(2) In case of Error No. 5 and No. 7
Calibration is continued. Unless another calibration
error occurs, calibration is carried out to the end, the
Measurement screen returns.
Error log file
If error occurs, the history is saved in an error log file. The error log file exists in the maintenance
mode.
Error log screen
Errors that occurred
* Up to 14 errors can be saved in the error history; the oldest error will be deleted one by one every
time a new occurs.
* If the power supply is turned OFF, the contents in the error log file will not be lost.
Deletion of error history
Press the
the
key will clear the error history.
INZ-TN2ZSVS-E
key on the above screen, and the "Error Log Clear" will be inverted. Further pressing
ENT
ENT
- 81 -
ESC
Date and time
when an error occurred
Channel with which
the error occurred
New
Old