4.1.3 Setting gas change time
Change gas is automatically fed when calibration is completed during measurement.
The duration of feed can be set as follows.
Description of gas change
• Gas change : After calibration is completed during measurement, sample gas is fed for a period
Setting range
• "30 to 300 sec" (in steps of 1 sec)
1) Move the cursor to <3. Setting about Gas change> using
the
key, and press the
The screen switches as shown at right.
2) Select a digit using the
using the
Press the
The cursor moves to the most significant digit of the
setting.
The setting has now been completed.
INZ-TN2ZSVS-E
of time arbitrarily selected.
key.
key, and change the value
key.
key to register the setting.
Initial value
"60 sec"
- 31 -
ENT
ESC
,
ENT
,
,
Setting complete