Caution on piping
Be sure to observe the following precautions while installing
piping. Improper piping may result in gas leakage.
If the leaking gas contains a toxic component, serious accidents
may result. If it contains combustible gases, explosion or fire
may result.
• Connect pipes correctly referring to the instruction manual.
• Discharge the exhaust gas outdoors to prevent it from remaining
within doors.
• Relieve the exhaust gas from the analyzer to the atmospheric
pressure to prevent buildup of undesirable pressure to the ana-
lyzer. Otherwise piping within the analyzer may be discon-
nected, resulting in gas leakage.
• Use pipes and pressure reducing valves to which no oil/grease is
attached for piping. Otherwise, fire may result.
Caution on wiring
• Be sure to turn off all the power before installing wiring. Other-
wise electric shock may result.
• Be sure to perform class D grounding work. Otherwise, electric
shock or failure may result.
• Select a proper wiring material that satisfies the ratings of the
instrument. Otherwise, electric shock or fire may result.
• Be sure to connect a power supply of correct rating. Otherwise,
fire may result.
Caution on use
• If unusual smell or sound has been produced, immediately stop
the instrument. Any discharge produced may cause a fire.
• Leaving the analyzer unused for a long time or restarting it after
long-term suspension requires procedures different from normal
operation or suspension procedures. Be sure to follow the
instructions in each instruction manual. Otherwise, intended
performance may not be achieved, or accidents or injury may
result.
• Do not operate the analyzer for a long time with its door left
open. Otherwise, dust, foreign matter, etc. may stick on internal
walls, thereby causing faults.
- iii -