2.2.1 Piping procedure

Caution on piping
• Use "Teflon" gas tube, which is low in adsorptive activity.
Tilt the gas tube that connects the gas extractor and the analyzer unit by about 30° to prevent
drain from staying within the tube.
• The maximum allowable gas tube length is 20m.
To extend the tube length, use a separate external pump.

2.2.2 Connecting gas extractor

(1) Flexible gas extractor (option)
The flexible gas extractor consists of a probe handle and a probe cap.
Probe handle

2.2 Piping in the gas analyzer

Gas extractor
Gas tube
About 30°
Probe cap
- 14 -
Sample gas inlet
INZ-TN2ZSVS-E