Figure 49. Apci Probe Nozzle Assembly - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

Thermo Scientific
To clean the APCI nozzle
IMPORTANT
Always use LC/MS-grade methanol and LC/MS-grade water (see
Table 7
on
page
55).
1. Remove the nozzle from the APCI probe body (see
2. Check the condition of the O-rings on the APCI nozzle

Figure 49. APCI probe nozzle assembly

3. Clean the interior APCI components (excluding the ceramic heater) with a 50:50
solution of methanol/water and a lint-free swab.
4. Dry the components with nitrogen gas to make sure that all the solvent evaporates.
5. Reinstall any O-rings removed while cleaning.
6. Do one of the following:
• If you do not want to replace the APCI sample tube, follow the procedure
reassemble the APCI probe"
• If you want to replace the sample tube, follow the procedure
sample tube"
on
0.312 in. ID 0-ring
0.500 in. ID 0-ring
on
page
78.
page
76.
5
Ion Source Probe Maintenance
page
74).
(Figure
49).
"To remove the APCI
LTQ Series Hardware Manual
Maintenance
"To
75