5
Maintenance
Electron Multiplier Replacement
120
LTQ Series Hardware Manual
4. Scroll down the Device list to display the electron multiplier voltages

Figure 83. Diagnostics dialog box showing the electron multiplier voltage parameters

5. For positive ion polarity mode, do the following:
a. Select the first multiplier device name.
b. In the Value box, type the mass spectrometer's device value listed in
Table 9. Electron multiplier device settings for positive ion polarity mode
Device name
Multiplier 1 high gain (V)
Multiplier 1 normal gain (V)
Multiplier 2 high gain (V)
Multiplier 2 normal gain (V)
c. Click Set.
d. Repeat
step 5a
through
Multiplier parameters
LXQ
–800
–800
–800
–800
step 5c
to set the remaining multiplier devices.
(Figure
83).
Table
9.
Device value
LTQ XL
Velos Pro
–900
–1200
–800
–1300
–900
–1200
–800
–1300