Cleaning The Mp00 Rf Lens Assembly Components; Figure 66. Mp00 Rf Lens Assembly (Ltq Xl) - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

5
Maintenance
MP00 RF Lens Maintenance

Figure 66. MP00 rf lens assembly (LTQ XL)

Lens L0

Cleaning the MP00 RF Lens Assembly Components

96
LTQ Series Hardware Manual
2.737 in. ID O-ring
Ball plungers (3x)
To clean the MP00 rf lens and lens L0
IMPORTANT
Always use LC/MS-grade methanol and LC/MS-grade water (see
Table 7
on
page
55).
1. Using a magnification device, inspect the components for any lint, particulates, and
sample buildup or coatings.
2. For 10 minutes, sonicate the components in either a 50:50 solution of methanol/water or
a 1% solution of Liquinox in water.
3. Using the 6000 grit MICRO-MESH polishing swabs, clean the bore in each lens.
4. Rinse the components thoroughly with water.
5. Sonicate the components in water for 10 minutes.
6. Sonicate the components in methanol for 10 minutes.
Ball plungers (6x)
Outer cage
MP00 rf lens
Contact ring
Threaded guide post (2x)