Figure 64. Mp00 Rf Lens Assembly Removal - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

5
Maintenance
MP00 RF Lens Maintenance
94
LTQ Series Hardware Manual

Figure 64. MP00 rf lens assembly removal

Vacuum
manifold
3.25 in. ID
O-ring
Hex socket-head
screws
6. Using the 5/32 in. hex ball driver, install two of the removed hex socket-head screws into
the two ejection screw holes
manifold.
7. Carefully remove the MP00 rf lens assembly and place it on the clean, lint-free work
surface.
8. Remove the two hex socket-head screws from the ejection screw holes.
Ejection screw holes
Outer cage
Ion source interface
(inside cage)
(Figure
64) to loosen the outer cage from the vacuum
Connector inside the
vacuum manifold
(not shown)