Maintenance
Electron Multiplier Replacement
118
LTQ Series Hardware Manual
To assemble the new electron multipliers
CAUTION Be careful not to damage the surface of the electron multiplier shield
(inside surface). This shield has been electropolished to prevent field emission.
1. Place the new electron multipliers on the work surface with the flat sides facing each other
(Figure
80).

Figure 80. Placement of screws for the left and right electron multipliers (top view)

2. For each multiplier, attach the HV cable to the HV connector that faces you.
The other HV connector is not used.
To replace the electron multiplier assembly
1. Place the new electron multipliers on the top cover so that the flat sides, if provided, face
each other.
2. Insert the HV connectors into the HV feedthroughs
Figure 81. Placement of the electron multipliers and shield retainers on the top cover plate
Use these screw holes.
High voltage connections to feedthroughs
Use these screw holes.
(Figure 80
on
page
118).
Electron multiplier
shield retainers