Diagnostics
Running General Diagnostic Tests
126
LTQ Series Hardware Manual
To run the general diagnostic tests
1. Open the Tune Plus window (see
2. Choose Diagnostics > Diagnostics to open the Diagnostics dialog box.
3. Click Tests at the top of the left pane to display the list of diagnostic tests

Figure 88. Diagnostics dialog box showing the general test page

4. Select one of the following options:
• To test all of the electronic subsystems (that is, the vacuum system, power supplies,
lenses, ion detection system, and rf voltage electronics), under General, select All, and
under What to Test, select the All Tests check box.
• To test an individual subsystem, click the item corresponding to that subsystem and
select the appropriate check box.
5. Click Start.
Testing starts and the Testing text box displays a chronological log of all diagnostic tests.
After tests are completed for a specific subsystem, the Result column displays either Pass
or Fail. If the diagnostics indicates a problem, contact your local Thermo Fisher Scientific
field service engineer for assistance.
page
34).
(Figure
88).