Functional Description
Mass Spectrometer

Power Panel

14
LTQ Series Hardware Manual
The power panel is located on the right side of the mass spectrometer.
show the power panels for the LXQ, LTQ XL, and Velos Pro mass spectrometers.
Figure 8.
Power panel for the LXQ
Reset button
Peripheral Control
Start
Ready
In
Out
Ethernet
100 Base T
Reset
Electronics Normal
1V
MAX
Analog Input
Service Mode
Electronics
service switch
Figure 9.
Power panel for the LTQ XL and Velos Pro
Reset button
Electronics
service switch
The mass spectrometer receives line power at 230 Vac ± 10%, 15 A (for the LTQ XL and
Velos Pro) or 10 A (for the LXQ), 50/60 Hz through the right-side power panel (see
Figure
8). The power panel provides system power control, a contact closure interface
(Peripheral Control), an Ethernet 100Base-T connection from the mass spectrometer to the
data system computer, and a system Reset button. The power panel accepts line power, filters
it, and provides it to various components of the mass spectrometer.
Start
Out
Main Power
10V
MAX
Figure 8
Power In
230 V AC
10 A
On
Mech. Pumps
Off
Main Power switch
Main Power switch
and
Figure 9
Thermo Scientific