Ion Source Probe Maintenance - Thermo Scientific LTQ Series Hardware Manual

Table of Contents

Advertisement

5
Maintenance

Ion Source Probe Maintenance

Ion Source Probe Maintenance
56
LTQ Series Hardware Manual
Table 7. Tools, equipment, and chemicals (Sheet 2 of 2)
Description
Water, LC/MS-grade
Water, tap
a
Provided in the HESI-II Probe Kit
b
Provided in the MS Accessory Kit
CAUTION AVOID EXPOSURE TO POTENTIALLY HARMFUL MATERIALS
By law, producers and suppliers of chemical compounds are required to provide their
customers with the most current health and safety information in the form of Material
Safety Data Sheets (MSDSs). The MSDSs describe the chemicals and must be freely
available to lab personnel to examine at any time. MSDSs summarize information on the
hazard and toxicity of specific chemical compounds. MSDSs also provide information on
the proper handling of compounds, first aid for accidental exposure, and procedures for
the remedy of spills or leaks.
Read the MSDS for each chemical you use. Store and handle all chemicals in accordance
with standard safety procedures. Always wear protective gloves and safety glasses when you
use solvents or corrosives. Also, contain waste streams, use proper ventilation, and dispose
of all laboratory reagents according to the directions in the MSDS.
This section contains information for maintaining the ion source probes:
"ESI Probe Maintenance"
"HESI-II Probe Maintenance"
"APCI Probe Maintenance"
You need the following tools and supplies (see also
Tools
Fused-silica cutting tool
Magnification device
7/64 in. hex driver
5/16 in. open-end wrench
3/8 in. open-end wrench
1/2 in. open-end wrench
on
page 57
on
page 69
on
page 72
Table 7
Supplies
Chamois-tipped swabs
Gloves, lint-free and powder-free
Kimwipes or lint-free industrial tissues
Methanol, LC/MS-grade
Nitrogen gas
Water, LC/MS-grade
Part number
Fisher Scientific
P/N W6-1
on
page
55).
Thermo Scientific