| 94 |
5 Service Interfaces
| CARDIOHELP System |
2
Start the service tool software (⇨ "Start Service Tool Software", page 85).
3
Switch to the "Special" tab.
4
Click on the
▶
The "Delete Configuration" warning is displayed.
5
Click on the
▶
If you do not want to delete the clinic configuration, click on the
▶
The warning "Clinical Configuration deleted" is displayed.
6
Click on the
5.3.9 Country Settings
WARNING!
The country settings are mandatory for USA.
Activate country settings
1
Activate the service interface (⇨ "Activate the Service Interface", page 79).
2
Start the service tool software (⇨ "Start Service Tool Software", page 85).
3
Switch to the "Special" tab.
4
Click on the
▶
The "Country Settings" pop-up screen is displayed.
5
Select the necessary check box.
▶
Mark the
▶
Mark the
6
Click on the
▶
The message "Country setting activated!" is displayed.
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
[Settings]
button.
[Yes]
button if you wish to continue.
[OK]
button.
[Country settings]
button.
[United states of America]
[Rest of the world]
check box for all other countries.
[Save]
button.
check box if the device will be used in USA.
[Settings]
symbol
"Delete Configuration"
[No]
button.
"Clinical Configuration deleted" warning
[Country settings]
symbol
Country settings
Country settings activated