| 52 |
3 Inspection
| CARDIOHELP System |
▶
Pump stops.
▶
Acoustic alarm sounds.
▶
Status bar turns red and displays the appropriate message.
u
Bubble detection flow/bubble sensor - passed/failed.
9
Connect the flow/bubble sensor to the HLS test circuit.
10 Deactivate the
11 Touch the
12 Touch the
▶
Speed regulates to 2500 rpm.
13 Set the speed to 0 rpm.
14 Connect the bubble sensor, if available, to the air bubble tube and repeat all steps.
CAUTION!
For the different tubing sizes 1/4" and 3/8" use the specific size of the bubble sensor.
u
Bubble detection bubble sensor - passed/failed/N/A.
Level Detection
1
Attach the level sensor to the level sensor pad.
▶
A blue tick is displayed to indicate the external level sensor.
NOTE
The "Level detection" intervention is only available in the "OR" application.
2
Touch the
3
Touch the
4
Touch the
5
Set the speed to 2000 rpm.
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
[Bubble monitoring]
[RESET]
symbol.
[Confirm]
symbol.
[Level monitoring]
symbol.
[Interventions]
symbol.
[Confirm]
symbol.
function.
Bubble alarm screen
Connected level sensor
Level monitoring
Level intervention screen
Need help?
Do you have a question about the MAQUET CARDIOHELP and is the answer not in the manual?