Saving The Serial Number; Delete Clinic Configuration - Save Factory Settings - Getinge MAQUET CARDIOHELP Service Manual

Heart-lung support system
Hide thumbs Also See for MAQUET CARDIOHELP:
Table of Contents

Advertisement

5
Choose the battery change date.
Normally the current date.
Except the battery change date must be restored.
6
Click on the
[Save]
The message "Set Battery Change Date" is displayed.
u
The battery change is saved in the CARDIOHELP.

5.3.7 Saving the Serial Number

CAUTION!
After replacing the HMI PCB, the serial number of the device must be saved on the new HMI PCB.
1
Activate the service interface (⇨ "Activate the Service Interface", page 79).
2
Start the service tool software (⇨ "Start Service Tool Software", page 85).
3
Switch to the "Special" tab.
4
Click on the
[Save Serial No]
The "Set Serial Number" screen is displayed.
5
Enter the device serial number of the rating plate.
6
Click on the
[Save]
The serial number is saved on the HMI PCB.
7
Click on the
[Close]
u
Device serial number is saved on the HMI PCB.

5.3.8 Delete Clinic Configuration - Save Factory Settings

1
Activate the service interface (⇨ "Activate the Service Interface", page 79).
button.
button.
button.
button.
| CARDIOHELP System |
5 Service Interfaces
"Saving the battery change"
"Set Batterie Change Date"
[Save Serial No]
symbol
"Set Serial Number" screen
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
| 93 |