2
Follow the instructions on the screen.
▶
A ping is sent.
Check Self-Test Status
1
Click on the
[Self-test status]
2
Follow the instructions on the screen.
Check Night Mode
1
Click on the
[Night Design]
2
Follow the instructions on the screen.
Check Alarm Outlets
1
Click on the
[Alarm Output]
2
Follow the instructions on the screen.
▶
The checking device K is required for this test.
Check CAN External Drive
1
Click on the
[CAN ext. Drive]
2
Follow the instructions on the screen.
▶
The checking device K is required for this test.
Check CAN External Clamp
1
Click on the
[CAN ext. Clamp]
2
Follow the instructions on the screen.
▶
The checking device K is required for this test.
Check Speaker
1
Click on the
[Speaker]
2
Follow the instructions on the screen.
Calibrate Sensors
1
Click on the
[Sensor calibration]
2
Follow the instructions on the screen.
Reset to Normal Mode
1
Make sure that all tests are passed.
button.
button.
button.
button.
button.
button.
button.
| CARDIOHELP System |
3 Inspection
Ethernet message
Self-test status message
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
| 57 |