13Battery; 14Speaker - Getinge MAQUET CARDIOHELP Service Manual

Heart-lung support system
Hide thumbs Also See for MAQUET CARDIOHELP:
Table of Contents

Advertisement

| 28 |
2 System Components
2.11.13 Battery
WARNING!
Do not mix old with new battery types.
Old battery type
New battery type
Main function
n
Power supply if mains power fails
– Lithium-ion battery
Technical data
n
Old battery type (CGR18650CG / M350-0149-BA0): Charging time 5 h max.
n
New battery type (NCR18650A / M350-0149-C00): Charging time 7h max.
Specification
Voltages
Output voltage
Storage capacity
Charging voltage
Charging current
2.11.14 Speaker
Speaker
Main function
n
Acoustic output device
Additional function
n
Internally sealed by a membrane
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
| CARDIOHELP System |
10.8 V DC
Old battery type: 6.75 Ah, 72.9 Wh
New battery type: 9.21 Ah, 99.47 Wh
12.45 V DC
max. 3 A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MAQUET CARDIOHELP and is the answer not in the manual?

Table of Contents

Save PDF