The following maintenance functions are summarized in this tab.
n
Creating the CARDIOHELP status report (⇨ "Create Status Report", page 61).
n
Analyzing service events and device errors (service pool) (⇨ "Analyzing Events and Errors", page 87).
– Events and device errors in the service pool are e.g., technical events
– Up to 260,000 errors messages can be saved
– Pool can only be deleted when all errors have been read out
n
Analyzing and deleting events and user actions (user pool)
(⇨ "Analyzing Events and Errors", page 87).
– Events and user actions in the user pool are e.g., user changes, switching on/off, changing limit
values...
– Up to 260,000 error messages can be saved
– Pool can only be deleted when all errors have been read out
n
Saving the service date (⇨ "Save Next Service Date", page 60).
n
Data export of the service actions carried out as a txt file for the service protocol (⇨ "Create Attachment
for Service Protocol", page 62).
n
Opening saved data files (txt) (⇨ "Analyzing Events and Errors", page 87).
5.2.5 "Special" Tab
"Special" tab
The following special functions are summarized in this tab.
n
Licensing (⇨ "Licensing", page 90).
n
Teaching the venous probe (⇨ "Teaching the Venous Probe", page 91).
n
Saving the battery change (⇨ "Saving the Battery Change", page 92).
n
Saving the serial number (⇨ "Saving the Serial Number", page 93).
n
Delete clinic configuration - save factory settings (⇨ "Delete Clinic Configuration - Save Factory
Settings", page 93).
n
Country settings (⇨ "Country Settings", page 94).
| CARDIOHELP System |
5 Service Interfaces
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
| 83 |
Need help?