3 Inspection
| CARDIOHELP System |
2
Click on the
3
Follow the instructions on the screen.
3.6.4 External Temperature Sensors
NOTE
These tests check the input signal of the sensor connections of the CARDIOHELP.
The measuring accuracy of the external temperature sensors is not tested.
Please use the MECC mode.
CAUTION!
Disconnect disposable connection cable!
If valid values of an integrated sensor are available, the value of the external sensor T
CAUTION!
Disconnect venous probe connection cable!
If valid values of venous probe are available, the values of the external sensor T
1
Touch the
2
Connect the connectors T
CARDIOHELP.
3
Set the T
the service protocol.
4
Read the individual T
service protocol.
u
External temperature sensor T
u
Repeat step 3 ... 4 for T
5
Remove the connectors T
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
[Normal mode]
button.
[Parameter list]
symbol.
and T
Art
values on the sensor tester "T External" according to
Art
values from the touchscreen of the CARDIOHELP and record them in the
Art
- passed/failed.
Art
. External temperature sensor T
Ven
and T
Art
of the sensor tester with the
Ven
from the CARDIOHELP.
Ven
Function test mode message Exit
Art
is ignored.
Ven
[Parameter list]
symbol
Connectors T
and T
Art
Sensor tester
- passed/failed.
Ven
is ignored.
Ven