6 Repair
| CARDIOHELP System |
19 Assemble the battery housing (⇨ "Assemble Battery Housing", page 117).
20 Close the device (⇨ "Closing the Device", page 110).
21 Carry out a complete inspection (⇨ "Inspection", page 33).
u
Power supply unit replaced.
u
Inspection after repair has been carried out.
6.17 Replacing Rotary Encoder
Requirements
Service training part 2
CAUTION!
ESD-sensitive components
To prevent damage to ESD-sensitive components, an ESD-protected environment must be set up.
WARNING!
To finish the task, the following step must be carried out in order to ensure error-free use.
n
Carrying out the inspection (⇨ "Inspection", page 33).
1
Remove the rotary knob (⇨ "Replacing the Rotary Knob", page 102).
2
Open the device (⇨ "Opening the Device", page 108).
3
Remove the battery housing (⇨ "Removing Battery Housing",
page 116).
4
Remove the backlight inverter (⇨ "Replacing Backlight Inverter", page 118).
5
Remove HMI PCB (⇨ "Replacing HMI PCB", page 119).
6
Unscrew the hex nut.
7
Remove the rotary encoder.
8
Remove silicon and clean the surface.
9
Replace the rotary encoder.
Service Manual | 3.3 | EN | 05
Copyright Maquet Cardiopulmonary GmbH
Rotary encoder
Battery housing
Rotary encoder