Seal wash pump was missing when tried to turn on

Error ID: 29223
The seal wash pump has not been detected (while being configured or detected
before)
Probable cause
1
2
3

Valve hardware overcurrent (MCGV)

Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause
1
2
3

Invalid degasser pressure signal

Error ID: 29253
The degasser pressure signal is invalid.
Probable cause
1
2
1260 Infinity II Flexible Pump User Manual
Defect cable connection to seal wash pump.
Defect seal wash pump motor.
Defective main board.
Cable defect.
Valve defect
Defective main board.
Degasser might be disconnected
Pressure sensor might be defective
Error Information
Pump Error Messages
Suggested actions
Check cable connection.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Replace MCGV.
Replace MCGV.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.