Error Information
Pump Error Messages

Target pressure not reached for binary pump degasser

Error ID: 22031
The target pressure of the binary pump degasser has not been reached within
the expected time.
• Parameter: Pressure in mbar
Probable cause
1
2

Degasser's pressure limit violation

Error ID: 22032
Pressure too far above the limit.
Probable cause
1
2
3
108
Condensation in degasser chamber due to
temperature fluctuation.
Degasser is defect.
Leak in degasser chamber or degasser
tubing.
Defect vacuum pump.
Degasser chamber empty or connected to
air.
Suggested actions
Equilibrate and restart module.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Block unused degasser channels.
Agilent 1290 Infinity Binary Pump User Manual