Module Error Messages; Solvent Zero Counter; Pressure Above Upper Limit - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

6
Error Information

Module Error Messages

Module Error Messages

Solvent Zero Counter

Error ID: 2055
The error message is triggered if the remaining volume in a solvent bottle falls
below the set limit.
Probable cause
1
2

Pressure Above Upper Limit

Error ID: 2014, 2500
The system pressure has exceeded the upper pressure limit.
Probable cause
1
2
3
4
84
Volume in bottle below specified volume.
Incorrect setting.
Upper pressure limit set too low.
Blockage in the flowpath (after the pressure
sensor).
Defective pressure sensor.
Defective main board.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Suggested actions
Refill bottles and reset solvent counters.
Make sure the set solvent volume matches the
actual bottle filling and set the shutoff limit to a
reasonable value (e.g. 100 mL for 1 L bottles)
Suggested actions
Ensure the upper pressure limit is set to a value
suitable for the analysis.
Check for blockage in the flow path. The following
components are particularly subject to blockage:
inline filter frit, needle (autosampler), seat capillary
(autosampler), sample loop (autosampler), column
frits and capillaries with small internal diameters (e.g.
50 µm ID).
Please contact your Agilent service representative.
Please contact your Agilent service representative.