6
Error Information
Module Error Messages
Degasser: limit not reached
Error ID: 2244, 8053
Degasser is not ready within 8 min after startup, because the pressure inside the
vacuum chamber exceeds 120 mbar.
Probable cause
1
2
3
SSV Protection circuit triggered
Error ID: 2049
The protection circuit of SSV has switched OFF the valve. During initialization, the
processor checks the presence of the SSV connector. If a defective valve or a
wrong connection is detected the error message is generated.
Probable cause
1
2
3
Agilent InfinityLab LC Series 1260 Infinity II Isocratic and Quaternary Pumps User Manual
Liquid in degasser tubing.
Leak in degasser tubing or chamber.
Degasser vacuum pump defect.
Wrong valve connected.
Defective Solvent Selection Valve.
Defective mainboard.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Check if the correct valve is connected to the
correct port.
Change the Solvent Selection Valve.
Please contact your Agilent service
representative.
110