Target pressure not reached for quaternary pump degasser

Error ID: 29221
The target pressure of the quaternary pump degasser has not been reached
within the expected time.
• Parameter: Pressure in mbar
Probable cause
1
2

Solvent counter exceeded limit

Error ID: 29146
The counter for the solvent volume has exceeded the limit, which has been set
in the user interface.
Probable cause
1
2

Waste counter limit exceeded

Error ID: 29147
The counter for the waste volume has exceeded the limit, which has been set
in the user interface.
• Parameter: None
Probable cause
1
2
1290 Infinity Quaternary Pump User Manual
Condensation in degasser chamber due to
temperature fluctuation.
Degasser is defect.
No solvent present.
Inappropriate setting.
The waste container is full.
Inappropriate setting for waste counter.
Error Information
Pump Error Messages
Suggested actions
Equilibrate and restart module.
Please contact your Agilent service
representative.
Suggested actions
Refill solvent bottle.
Check solvent counter setting in user interface.
Suggested actions
Empty waste container.
• Reset waste counter.
• Adjust waste counter limit.