General Information; Equipment Required To Install Pump Module; Tools Required To Install Pump Module - Agilent Technologies Division TriScroll 600 Series Replacement Manual

Dry scroll vacuum pump
Table of Contents

Advertisement

General Information

AgilentTriScroll 600 series pumps are designed to provide years of trouble-free service if maintenance
procedures and intervals are observed. Bearing grease replenishment and tip seal replacement is
recommended when the pump base pressure has risen to an unacceptably high level for your
application. Bearings, rotary seals and O-rings should also be replaced if the pump exhibits humming
or grinding noises from the bearings. Main bearing life may be shortened if your application requires
the pumping of high quantities of water vapor. Use of the bearing purge will keep this water from
impacting bearing life.

Equipment Required to Install Pump Module

Pump Module

Tools Required to Install Pump Module

4 MM Hex key
5 MM Hex key
6 MM Hex key
Included in Maintenance Tool Kit: #PTSS0600TK (page 2)
TriScroll 600 Dry Scroll Vacuum Pump
1

Advertisement

Table of Contents
loading

Table of Contents