Road maps to start an analysis with the nano pump

Road map to configure the nano pump

The road map below guides you through all parameters you may need to
configure your nano pump.

Road map to prepare the nano pump for an analysis

The road map below gives you an overview of the necessary steps to prepare
the nano pump before starting an analysis.
When the pump is used for the first time or has been turned off for a certain time, it is
N O T E
recommended to purge each channel at 2500 µl/min for 1 minute at least.
Before running the pump in micro mode, run it in normal mode until the system pressure is
stable.
18
Setup the
Setup the
Primary Flow
Column Flow
Ready Condition
page 21
page 21
Setup the
To configure the nano pump
Mixer Volume
page 20
Start
Here!
Setup the
Setup the
Minimum
Solvent
Stroke
Compressibility
page 22
page 22
Setup the
Solvent
Calibration
page 26
1200 Series Nano Pump User Manual