Manuals and User Guides for Siemens ADVIA 1800. We have
1
Siemens ADVIA 1800 manual available for free PDF download: Operator's Manual
Siemens ADVIA 1800 Operator's Manual (222 pages)
Chemistry system
Brand:
Siemens
| Category:
Measuring Instruments
| Size: 6.77 MB
Table of Contents
Table of Contents
3
1 System Overview
13
Operating Principle
13
Hardware Overview
14
Display and Power Panel
15
Analyzer Back View
15
External Circuit Ratings
16
External Circuits
16
Connected Equipment
16
Analyzer Top View
16
STT Rotate and SMP Pause Buttons
17
Pumps
17
Workstation (Front View)
18
PC (Front View)
18
PC (Back View)
19
Opening the Top Cover
19
Reagents
19
Reagent Mechanisms
19
Reagent Trays
20
Reagent Tray Operation
20
RGT Pause Button
21
Reagent Pumps
21
Reagent Container Types
22
Sampling and Analysis of Photometric Tests
22
Sample Aspiration
22
Sample Tray
22
SMP Pause
23
Sample Aspiration and Dilution Mechanisms
24
Dilution Pumps (DIP and DOP)
24
Sample Pumps
25
Sample Dilution
25
Dilution Tray Operation
25
Dilution Tray Mixer
26
Dilution Tray Mixer Operation
26
Dilution Tray Wash Mechanisms
26
DPP Probe Features
27
Sample Delivery to Reaction Tray
28
Sampling Mechanisms
28
Sample Probe
29
Analysis
29
Reaction Tray
29
Reaction Tray Operation
30
Reaction Tray Mixers
31
Reaction Tray Mixer Operation
31
Reaction Tray Wash Mechanisms
32
Reaction Washer Operation
32
Reaction Tank
34
Spectrophotometer
34
Universal Rack Handler - Functional Description (Optional)
36
Software Overview
38
Startup Window
38
Operation Panel
39
Buttons
39
Status and Message Boxes
39
Other Tasks Accessed from System(S) on the Operation Panel
39
Menu Panel
40
Other Tasks Accessed from System(S) on the Menu Panel
40
Description of Software Windows
41
Request Windows
41
Order Entry Window
41
Sample Log Window
42
Test Result Monitor Window
42
Sample Test Status Color Codes
43
Review/Edit Window
44
Reaction Monitor Window
45
Realtime Monitor Window
45
Test Select Window
45
Cup/Tube Assign Window
45
Print Report Window
45
Lab Results Converter (LRC)
46
Setup to Use LRC
46
Printing And/Or Generating a PDF from the LRC
46
Statistics Window
46
Correlation Window
46
Data Transmit Window
47
Calibration Windows
47
View Calibration Curve Window
47
Calibration/Rbl History Window
47
Sample Select Window
48
Calibration Setup Window
48
Maintenance Windows
49
System Startup/Shutdown Setting Window
49
User Maintenance Window
49
System Monitor Window
49
System Maintenance Monitor Window
49
Lamp Energy Monitor Window
49
ISE Operation Window
50
ISE Monitor Window
50
Manual Operation Window
50
On-Line Monitor Window
51
Error Report Window
51
Viewing Message Details
52
Controlling the Number of Messages Listed
52
Updating the Message List
52
Printing the Messages
52
Deleting the Messages
52
JEOL Maintenance Window
52
Liquid Level Sensor Monitor Window
53
LAS Monitor Window
53
Test Utilization Statistics Window
53
Reagent Windows
53
Reagent Inventory Window
53
Monitoring Reagent Positions, Number of Tests, Volumes, and Days Remaining
54
Notes
54
Executing a Barcode Scan
55
Deselecting a Reagent
55
Printing a Reagent Summary Report
56
Viewing Total Test Summary
56
Viewing Cal Interval Review
56
CTT Monitor Window
56
Reagent Container Settings Window
57
Active Test List Window
57
Reagent Information Window
57
Reagent Barcode Maintenance Window
57
QC Windows
58
ADVIA QC Window
58
Real-Time QC Window
58
Daily Precision Control Window
59
QC Cumulative Window
59
Control Data Setup Window
60
Sample Select Window
60
QC Sample Definition Window
60
Setup Windows
61
System Specification Settings Window
61
System Test List Window
62
Analysis Order Set Window
62
Analytical Parameters (Serum) Window
62
ISE Parameter Settings Window
63
Reflex Test Settings Window
63
Print Form Settings Window
63
Ctrl/Cal Sample Setup Window
63
Setting System Parameters Window
64
New Test Definition Window
64
Analytical Parameters (Chemistry) Window
64
Process Sequence Window
66
Ratio Parameters Window
66
Blank Reagent Settings Window
66
Contamination Settings Window
66
Online Settings Window
67
Alarm Buzzer Settings Window
67
User Code Settings Window
67
Other Tasks
67
Version Information Window
68
ADVIA O.p
68
Controller
68
Safety Message
68
Etc
69
Memo
69
Using Online Help
69
Viewing the Content
70
2 Operating the System
71
Start the System
71
Logging on
71
Starting the System
71
Performing a System-Assisted Startup (Start Set)
72
Checking the Analyzer
73
Set up of Daily Washes
77
Times Required to Perform Prime, Washes, and Cell Blank
78
Daily Operation
78
Using Workorders
78
Using Host Computer Workorders
78
Creating Workorders at the Analyzer
79
Creating Multiple Workorders
79
Deleting or Changing Test Selectivity for Multiple Workorders
80
Creating a Profile
81
Creating a Load List
81
Loading Patient Samples
82
Loading Patient Samples on the STT
82
Methods Introduction
83
Sample Collection and Preparation
83
Loading Patient Samples on the Optional Universal Rack Handler
85
Urgent Samples and Manual Reruns Are Run from the Onboard Sampler (STT)
87
Starting the Run
87
Running Calibrators and Reagent Baselines (RBL)
87
Running Control Samples
88
Monitoring Samples Loaded on the Optional Universal Rack Handler
89
Running an Interrupt (STAT) Sample
89
Monitoring an Analysis While It Is Running
90
Processing Status
90
Sample Information and System Status
90
Result Data Information
91
Using the Sample Log
91
Viewing the Sample Log Entries
91
Searching the Sample Log
92
Deleting a Specific Sample Log Entry
92
Deleting All Sample Log Entries
92
Printing a List of the Sample Log Entries
92
Exporting the Sample Log Entries
93
Reviewing the Calibration Results
93
Reviewing the Sample Results
93
Reporting Results
93
Real-Time Reporting
93
Batch Transmission to a Host Computer
94
Printing Sample Data
94
Batch Printing
95
Shutting down the ADVIA Chemistry System
95
Examples of Shutdown Settings
96
System Available During the Night
96
Automatic Startup from the Sleep State
96
Automatic Startup with Analyzer Power Kept on
96
Performing a System-Assisted Shutdown (Shutdown Set)
97
Emergency Stop, Shutdown, and Recovery Procedures
98
Stopping Operation after the Analyzer Is Started
98
Managing an Expected Power Outage
98
Responding During a Power Failure While Electric Power Is Still off
98
To Resume Operations after Power Is Restored
98
Additional Operating Instructions
99
Switching Process Sequence and System Test List Numbers
99
Loading the Reagent Trays
99
Updating the Cumulative QC
99
Running Automatic Calibration
100
Using Multiple Reagent Containers for the same Test
100
Auto-Calibration and Auto-QC Situations that Need Corrective Steps
101
Auto Calibration with System-Related Errors
101
Auto-Calibration with Lot Change
101
Reagent Pack Switch in MID-Calibration
101
3 Calibration
103
Calibration Overview
103
When to Calibrate
103
Setting up the Calibration
103
Entering Absolute or Single-Point (STD) Calibration Methods
103
Entering Multi-Standard Calibration Methods
104
How to Calibrate a Multi-Standard Method
105
Running Calibration
105
Reviewing the Calibration
106
Using the Calibration/Rbl History Window
106
4 Quality Control
109
Quality Control Overview
109
When to Run Control Samples
109
Setting up Quality Control
109
Running the Control Samples
109
ADVIA QC Window
110
Reviewing the Control Results at Different Windows
110
5 Maintenance
111
Maintenance Schedule
111
Analytical Parameter Audit Trail
114
Daily Maintenance
115
Inspecting and Cleaning the Probes
115
The Probe Cleaning Tool Kit
117
Kit Contents
117
Procedure
118
Inspecting and Cleaning the Mixing Rods and Mixer Wash Cups
119
Checking Reagents and System Solutions
120
Monitoring Reagent Tray Temperature
120
Inspecting and Cleaning the Reaction (WUD) and Dilution (DWUD) Cuvette Washers
121
Inspecting and Cleaning the Cuvette Splash Covers
123
Inspecting and Cleaning the Probe Wash Cups
124
Inspecting the Pumps
125
Performing the Startup Wash (WASH3)
127
Performing the Shutdown Wash
128
Performing Additional ISE Electrode Washes
129
Recording ISE Slopes
130
Weekly Maintenance
131
Performing the Weekly Wash
131
Checking and Replenishing the Lamp Coolant
132
Checking Lamp Energy
133
Reading Lamp Energy Data
135
Measuring Cuvette Blanks
135
Cell Blank Measurement Results Are Summarized as Follows
136
Cleaning the Analyzer and Rack Handler Exterior Panels
137
Monthly Maintenance
138
Cleaning the Turntable Interiors (STT/CTT and RTT)
138
Cleaning the Inside of the STT/CTT Housing
138
Cleaning the Inside of the Reagent Tray Refrigerated Housing
139
Cleaning or Replacing the Wash Solution Reagent Containers (53 - 56)
140
Cuvette Wash and Cuvette Conditioner Usage
140
Cleaning and Replenishing the Dilution Bottle
141
Cleaning and Replenishing the Cuvette Wash Bottle
142
Cleaning the Chiller Filter
143
Cleaning the Large Water Pump (LWP) Filter
144
Every 2 Months Maintenance
146
Cleaning the Dilution Tray Cuvettes
146
Cleaning and Replenishing the Cuvette Conditioner Bottle
147
Every 3 Months Maintenance
149
Replacing the Lamp
149
Washing the Electrodes Lines
151
Every 4 Months Maintenance
153
Cleaning the Ancillary Reagent Bottle Filters
153
Cleaning the Pure-Water Bottle Filter
154
Replacing the Reaction and Dilution Cuvettes
155
As Required Maintenance
158
Backing up System Files
158
Restoring System Files
159
Replacing the SPP, RPP1, and RPP2 Probes
160
Removing a Probe
160
Install a New Probe
162
Replacing DPP Probes Equipped with Crash Detection
163
Removing the DPP Probe Cover
163
Removing the Probe from the Probe Arm
164
Install the New DPP Probe
165
Aligning the DPP Probe
166
Replenishing the RRV (Reaction) Bath Oil Bottle
166
Preventive Cleaning of the Wash Station Lines
168
Washing All the ISE Lines
172
Washing the Lines
173
Rinsing the Lines
173
Priming and Initializing the ISE Module
174
Replacing ISE Electrodes
174
Removing Electrodes
174
Installing Electrodes
175
Priming the Ises
175
Calibrating the Ises
176
Storing the Reference Electrode
176
Conditioning the Na and K Electrodes
177
Cleaning the Dilution Bowl and Waste-Drain Nozzle
178
Cleaning the Dilution Bowl
178
Cleaning the Waste Drain Nozzle
179
Maintaining the ISE Unit after the Dilution Bowl and Waste-Drain Nozzle Are Clean
180
6 Troubleshooting
181
7 File Management
183
Saving the Error Log to a Test File
183
Capturing the Error Report
183
Saving the Test Data from the User Maintenance Window
184
Saving Test Data
184
Saving Reaction Data in CSV Format
185
Saving a Data Archive from User Maintenance Window
186
Viewing Data that was Previously Archived
187
8 System Setup
189
Connecting the ADVIA Chemistry System and Your Host Computer
189
Setting the System Monitor
189
During Operation System Settings Area
190
Pre-Operation Settings
190
Clot Detection Settings
190
Setting up the ISE
190
Activating the ISE
190
Setting up ISE Parameters
191
ISE Calibration
191
Checking ISE Calibration
191
Configuring Reruns
193
Automatically Request Reruns for Flagged Samples
193
Processing the Rerun Automatically
193
Manually Requesting Rerun Using Review/Edit
193
Appendix A - Safety Information
195
Warning and Hazard Statements
195
Biohazard Warning
195
Electrical Warning
195
Laser Warning
195
Household Bleach Warning
196
Ancillary and Method Reagents Warnings
196
Regulatory Compliance
197
Laser Information
197
Appendix B - Warranty and Support Information
199
Limited Instrument Warranty and Service Delivery Policy
199
Warranty Period
199
Additional Service Period
199
Service Calls
200
Service During Normal Hours
200
Extent of a Service Call
200
Service Outside Normal Hours
200
Replacement of Parts
200
Warranty and Service Exclusions
200
Design Changes and Retrofitting of Instruments
202
OSHA Requirements
202
Software License
202
ADVIA Contact Information
203
Siemens Offices Worldwide
203
Appendix C - Customer Replaceable Parts
205
Appendix D - Specifications
207
Specifications - All
207
Appendix E - Symbols
213
Explanations of Symbols Associated with the ADVIA 1800 System
213
Warning and Caution Symbols
213
System Operation Symbols
214
System Rating Label Symbols
215
Hardware Component Symbols
216
Ancillary Reagent Symbols
216
Connector Symbols
217
System Packaging Label Symbols
218
Appendix F - Flags
219
Alarm (Result) Flags
219
Advertisement
Advertisement
Related Products
Siemens SENTRON ATC6300
Siemens Milltronics A-300
Siemens ARDOCELL PA 40 AF 3/L
Siemens ARDOCELL PA 40 AF 21
Siemens ARDOCELL PA 40 AF 5
Siemens ARDOCELL PA 40 AF 22
Siemens ARDOCELL PA 43 AF 10
Siemens ARDOCELL PA 43 AF 13/C
Siemens ARDOCELL PA 43 AF 8
Siemens ARDOCELL PA 43 AF 12
Siemens Categories
Industrial Equipment
Controller
Control Unit
Oven
Washer
More Siemens Manuals
Login
Sign In
OR
Sign in with Facebook
Sign in with Google
Upload manual
Upload from disk
Upload from URL