TELEDYNE Model 9110AH NO

4.2.8 Ozone Generator

Because of the instability of ozone, it is necessary to generate this gas inside the analyzer. The
ozone generation module consists of a switching power supply that drives a high voltage
transformer and silent discharge tube. The generator's dry air supply uses a membrane drier to
supply air with a dew point of O o C or less. The exhaust side of the membrane is connected to
the vacuum manifold at the rear of the instrument. A complete description and service
requirements for this module can be found in Section 9.3.6.
Normal room air contains enough water vapor to damage the generator and components
downstream, therefore the ozone GENERATOR MAY NOT TURN ON IMMEDIATELY
AFTER POWER UP. The delay is built into the instrument to allow the dryer to start operating
and purge the system with dry air. Table 4-3 details the conditions for turning on the ozone
generator.
Table 4-3: Ozone Generator Start-up Timing
Time Since Last Power-up
< 1 hour
> 1 hour
4.2.9 NO

- NO Converter

2
The converter is a length of 1/8" tubing heated to 700 C. The converter's function is to reduce
nitrogen dioxide (NO
by the computer. To accurately measure converter efficiency, there should be oxygen present in
the NO
calibration gas.
2
For the converter to operate properly there should be a few
Analyzer Operator Manual, 01620, Rev. F
X
Ozone Gen State
ON at power-up
OFF at power-up
) to nitric oxide (NO). The temperature control for this module is done
2
NOTE
percent oxygen present in the sample stream.
4-10
Program Action
Gen ON immediately after power-up.
Wait 30 min, then turn gen ON.