Calibration Date Features

At the end of warm-up and before the First Responder CG
is ready for measuring, the instrument will indicate on the
display, how many days remain until calibration is required,
as shown in Figure 3-4:
Calibration Due
(Alternate flash)
in 25 days
Figure 3-4 Days until Calibration Required
If calibration is overdue, the instrument will automatically
switch OFF.
The pre-set re-calibration interval is 30 days.
This period can be altered as an option, however, you should
ensure that the instrument is always within its calibration
period prior to use.

Switching OFF

A double press of
(Button 1) switches the instrument
OFF.

Switching Pump OFF / ON

A single press of
(Button 1) switches the instrument
pump OFF. A further press of Button 1 switches the pump
ON again.

Manual Datalog

Manual datalogging is enabled in this mode.
To perform a manual datalog, a double press of
(Button
3) is required.