CAUTION
Even with alarms set, if "Not detected" has been set in the above alarm detection, no alarm is
issued. Be sure to set "Detected" in the above alarm detection if you use alarm features.
8.3.4
Default Values
When the analyzer is delivered, or if data are initialized, the default alarm set values are as
shown in Table 8.8.
Table 8.8
Oxygen concentration high-high alarm setpoint
Oxygen concentration high alarm setpoint
Oxygen concentration low alarm setpoint
Oxygen concentration low-low alarm setpoint
Humidity high-high alarm setpoint
Humidity high alarm setpoint
Humidity low alarm setpoint
Humidity low-low alarm setpoint
Mixing ratio high-high alarm setpoint
Mixing ratio high alarm setpoint
Mixing ratio low alarm setpoint
Mixing ratio low-low alarm setpoint
Oxygen concentration alarm hysteresis
Humidity alarm hysteresis
Mixing ratio alarm hysteresis
Delayed alarm action
High-high alarm detection
High alarm detection
Low alarm detection
Low-low alarm detection
8.4
Output Contact Setup
8.4.1
Output Contact
Mechanical relays provide contact outputs. Be sure to observe relay contact ratings. (For details,
see Section 2.1, "General Specifications.") The following sets forth the operation mode of each
contact output. Output contact 1 you can select open or closed contact when the contact is
"operated". For output contact 2, contact is closed. The relay for output contact 1 is energized
when its contacts are closed and vice versa. Accordingly, when no power is supplied to the
equipment, those contacts remain open. In addition, the relay for output contact 2 is energized
when the corresponding contact is open and de-energized when that contact is closed.
Alarm Setting Default Values
Set item
<8. Detailed Data Setting>
Set value
100% O
2
100% O
2
0% O
2
0% O
2
100% H
O
2
100% H
O
2
0% H
O
2
0% H
O
2
1 kg/kg
1 kg/kg
0 kg/kg
0 kg/kg
0.1% O
2
0.1% H
O
2
0.001 kg/kg
3 seconds
Not detected
Not detected
Not detected
Not detected
IM 11M12A01-04E
8-10