System Power-Up - GE Optima CT680 Series Installation Manual

Table of Contents

Advertisement

GE C
OMPANY
D
5472001-1EN, R
IRECTION
EVISION
4.) Use a 0-750 AC voltmeter of 3/4% accuracy to measure the line-to-line voltages at L1, L2, &
WARNING
THIS PROCEDURE MEASURES POTENTIALLY HAZARDOUS VOLTAGES. USE AND
FOLLOW LOCKOUT/TAGOUT PROCEDURES.
TAG
&
LOCKOUT
Signed
Date
No Load
Line to Line Voltages
Nominal
480V*
460V
440V
420V
400V
380V
240V**
220V**
200V**
* Factory Default
** 2326492-3 PDU only
Table 4-1 PDU Line Tap Connections
Record system voltages here:
Phase A: _______ Phase B: _______ Phase C: _________
6.7

System Power-Up

CAUTION
Verify all personnel have cleared the system before you turn on wall power.
1.) Turn ON the A1 breaker panel.
Note:
Do not stand in front of the main disconnect to turn on power.
2.) Turn ON all system power switches and breakers (PDU, gantry, table, console).
Page 162
6
L3.
-
Verify the highest line-to-line voltage does not exceed 1.02 times the lowest voltage.
-
Example: If the lowest voltage equals 474, the highest voltage should not exceed 474 x
1.02 = 483.5 volts.
Maximum Range
(10%)
432 to 528*
414 to 506
396 to 484
378 to 462
360 to 440
342 to 418
216 to 264**
198 to 242**
180 to 220**
-
All PDU breakers
-
Make sure that the on/off button (on the front PDU panel) is ON for console power.
Section 6.0 - Electrical Power On & Ground Checks
O
CT680 S
PTIMA
ERIES AND
Tap Connections
(All 3 phases must have same the
configuration)
Phase A
Phase B
Connection
Connection
3-4*
3-4*
3-5
3-5
3-6
3-6
2-4
2-4
2-5
2-5
2-6
2-6
1-4**
1-4**
1-5**
1-5**
1-6**
1-6**
O
CT670 I
M
PTIMA
NSTALLATION
Phase C
Connection
3-4*
3-5
3-6
2-4
2-5
2-6
1-4**
1-5**
1-6**
ANUAL