Table 2-16 Panel Circuit Breaker Descriptions
8.1.3

HVDC Connection

Note:
Refer to
Connect the internally shielded HVDC cable to TS2 on the standing panel. See
location of the connector and
shorten cables unless you have all of the appropriate tools and crimper to re-terminate. If short
cables are needed, have the PMI order the short cable set.
WARNING
Excess cable length cannot be stored under or behind the PDU. If cables are to be stored in the
cable tray, do not overfill. Consult the local electrician to determine the maximum fill rate for your
area.
Figure 2-31 HVDC Connection
Check box when complete.
6
DESCRIPTION
Circuit Protection (Axial Drive)
Full Winding Protection
CT Gantry Service Outlets
CT Gantry rotating loads
Table & CT Gantry Stationary Loads
Operator Console Load
VMR1, Control P.S Load
VMR2
Table 2-3, System Interconnect Cables on page
Figure 2-31
Standing Panel - HVDC SUPPLY
OUTPUT CONNETCTIONS
Install Shield Clamp tightly
around exposed Cable Shield
at lower left of PDU.
Chapter 2 - Power, Ground & Interconnect Cables
O
CT680 S
O
PTIMA
ERIES AND
PTIMA
100.
for details. Observe polarities and grounds. Do not cut or
CT670 I
M
NSTALLATION
ANUAL
Figure 2-29
for the
Page 127