Section 9.0 Table Installation (Gt1700V); Time And Personnel; Tools And Test Equipment; Procedures - GE Optima CT680 Series Installation Manual

Table of Contents

Advertisement

GE C
OMPANY
D
5472001-1EN, R
IRECTION
EVISION

Section 9.0 Table Installation (GT1700V)

Note:
For Lite Table Installation, refer to
9.1

Time and Personnel

Required Persons
2 (FE or mechanical supplier)
9.2

Tools and Test Equipment

9.3

Procedures

9.3.1

Draw Table Reference Lines

1.) Draw a reference line of 673 ± 6 mm (26.5" ± 0.25")position from Gantry Base on the floor as
Figure 1-25 Draw Reference Line
Page 54
6
Standard Install Tool Kit
3/4", 1-1/4", 1-1/2" and 1-5/8" sockets
8mm, 10mm, and 14mm hex socket bits
Laser Alignment kit
Johnson Professional 6" level
Johnson Professional 4' level
Johnson Professional 2' level
shown in the
Figure
1-25. This line should be parallel to the gantry. In a later section, you will
move the table against the 673 mm (26.5") mark.
Draw this line on the floor.
Section 9.0 - Table Installation (GT1700V)
O
CT680 S
PTIMA
ERIES AND
Section
10.0.
Preliminary Reqs
Procedure
1.5 hours labor on-site
673 ± 6 mm (26.5" ± 0.25")
O
CT670 I
PTIMA
NSTALLATION
Finalization
Gantry Base
M