GE C
OMPANY
D
5472001-1EN, R
IRECTION
EVISION
6.) Verify that less than 1 ohm of resistance exists between the following connections:
Page 134
6
FROM
SIGNAL NAME
(COLOR)
PDU TS2-1
+HVDC
(Red)
PDU TS2-2
HVDC Ground
(Green/Yellow)
PDU TS2-3
-HVDC
(Black)
PDU Ground Bus HVDC shield
PDU TS3-1
Axial drive 440vac
(Black)
PDU TS3-2
Axial drive 440vac
(Red)
PDU TS3-3
Axial drive 440vac
(Orange)
PDU TS3-4
Axial drive ground
(Green/Yellow)
PDU Ground Bus Axial drive shield
PDU TS5-1
120vac Phase A
(Brown)
PDU TS5-2
120vac Neutral
(Light Blue)
PDU TS5-3
Ground
(Green/Yellow)
PDU TS5-4
120vac Phase A
(Black)
PDU TS5-5
120vac Phase B
(Red)
PDU TS5-6
120vac Phase C
(Orange)
PDU TS5-7
120vac Neutral
(Light Blue)
PDU TS5-8
Ground
(Green/Yellow)
Table 3-2 Resistance Verification Points
Section 1.0 - System Continuity (Mechanical Contractor)
O
CT680 S
O
PTIMA
ERIES AND
PTIMA
TO
Gantry HV Power Pan
TS1-1
Gantry Power Pan Chassis Check box when
Gantry HV Power Pan
TS1-2
Gantry HVDC cable shield Check box when
Gantry HV Power Pan
TS2-1
Gantry HV Power Pan
TS2-2
Gantry HV Power Pan
TS2-3
Gantry Power Pan Chassis Check box when
Gantry 440 VAC cable
shield
Console Power Plug:
BLU
W
BRN
G
GRN/YLW
Gantry LV Power Pan TS4-
1
Gantry LV Power Pan TS4-
2
Gantry LV Power Pan TS4-
3
Gantry LV Power Pan TS4-
4
Gantry Power Pan Chassis Check box when
CT670 I
M
NSTALLATION
ANUAL
Check box when
complete
complete
Check box when
complete
complete
Check box when
complete
Check box when
complete
Check box when
complete
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
complete

Hide quick links: