GE C
OMPANY
D
5472001-1EN, R
IRECTION
EVISION
6.) Verify that less than 1 ohm of resistance exists between the following connections:
Page 134
6
FROM
SIGNAL NAME
(COLOR)
PDU TS2-1
+HVDC
(Red)
PDU TS2-2
HVDC Ground
(Green/Yellow)
PDU TS2-3
-HVDC
(Black)
PDU Ground Bus HVDC shield
PDU TS3-1
Axial drive 440vac
(Black)
PDU TS3-2
Axial drive 440vac
(Red)
PDU TS3-3
Axial drive 440vac
(Orange)
PDU TS3-4
Axial drive ground
(Green/Yellow)
PDU Ground Bus Axial drive shield
PDU TS5-1
120vac Phase A
(Brown)
PDU TS5-2
120vac Neutral
(Light Blue)
PDU TS5-3
Ground
(Green/Yellow)
PDU TS5-4
120vac Phase A
(Black)
PDU TS5-5
120vac Phase B
(Red)
PDU TS5-6
120vac Phase C
(Orange)
PDU TS5-7
120vac Neutral
(Light Blue)
PDU TS5-8
Ground
(Green/Yellow)
Table 3-2 Resistance Verification Points
Section 1.0 - System Continuity (Mechanical Contractor)
O
CT680 S
O
PTIMA
ERIES AND
PTIMA
TO
Gantry HV Power Pan
TS1-1
Gantry Power Pan Chassis Check box when
Gantry HV Power Pan
TS1-2
Gantry HVDC cable shield Check box when
Gantry HV Power Pan
TS2-1
Gantry HV Power Pan
TS2-2
Gantry HV Power Pan
TS2-3
Gantry Power Pan Chassis Check box when
Gantry 440 VAC cable
shield
Console Power Plug:
BLU
W
BRN
G
GRN/YLW
Gantry LV Power Pan TS4-
1
Gantry LV Power Pan TS4-
2
Gantry LV Power Pan TS4-
3
Gantry LV Power Pan TS4-
4
Gantry Power Pan Chassis Check box when
CT670 I
M
NSTALLATION
ANUAL
Check box when
complete
complete
Check box when
complete
complete
Check box when
complete
Check box when
complete
Check box when
complete
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
Check box when
complete
complete