2.) Using a chalk line, mark a table center line on the floor along the laser light shining on the floor.
Reference line for Table 673 mm (26.5") Z-distance
Figure 1-26 Draw Reference Line for Table Center
3.) Recheck the table-to-gantry reference line for 673 ± 6 mm (26.5" ± 0.25") Z-distance. Refer to
Figure
4.) Turn off the laser but do not remove.
9.3.2
Table Prep and Set-up
SAFETY
CAUTION
Potential for Electric Shock.
Equipment is Energized.
Follow appropriate safety procedures when working with an energized system.
CAUTION
Potential for Injury.
Table will tip if not anchored on the dolly.
Make certain that Table is adequately secured to the dolly.
CAUTION
Potential for Injury.
Table on dolly length is 2.5m (98 in.).
Exercise caution when moving the table on the dolly.
PROCEDURE
CAUTION
Potential for Injury.
Table will tip if not anchored on the dolly.
Make certain that Table is adequately secured to the dolly.
6
Reference line for Table center
1-26.
Chapter 1 - Position Subsystems
O
CT680 S
O
PTIMA
ERIES AND
PTIMA
673 ± 6 mm (26.5" ± 0.25")
Gantry Base
CT670 I
M
NSTALLATION
ANUAL
Page 55