Section 3.0 Shim Installation

3.1

Time and Personnel

Required Persons
1 (FE or mechanical supplier)
3.2

Tools and Test Equipment

NOTICE
Understand and Follow All General Table Safety Procedures.
3.3

Preparation

Check head holder for a tight fit. If the head holder fit is loose, follow this procedure and shim for.
Introduction:
Figure 3-3 Axial Head Holder
Notes before Selecting Shim Thickness:
Page 138
6
Standard FE Tool Kit
Shim Kit
Axial head holder
Foot extender
Phantom holder
Some Axial Head Holders have a large free-play in the horizontal direction which could
potentially lead to motion and therefore image artifacts.
Installation of the 2327335 rubber shim kit can minimize this motion.
While selecting the best shim size, do not attach the rubber shim yet using the adhesive on the
back. It is best to use a piece of tape to hold on the shim in order to see if the size is correct.
Selecting a shim size that is too thick may result in:
-
Difficulty latching the head holder properly. The head holder must latch so that a patient
is not injured.
Section 3.0 - Shim Installation
O
CT680 S
PTIMA
ERIES AND
Preliminary
Procedure
Reqs
10 min.
15 minutes labor on-site 5 min.
O
CT670 I
M
PTIMA
NSTALLATION
Finalization
ANUAL

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima ct670Optima ct680 quantumOptima ct680 expertOptima ct680 professional

Table of Contents