Removing Table Shipping Dollies

11.10.1

Time and Personnel

(FE or mechanical supplier)
Required Persons
1 Electrician
11.10.2

Tools and Test Equipment

Standard Installation Tools Kit
11.10.3

Preparation

All table mechanical alignment procedures are completed.
The table is on the floor with at least one anchor in place.
11.10.4

Procedure

Refer to
1.) Remove the two, long side (stabilizer) rails using the quick disconnect pins. There is one pin
on each end of the bar.
2.) Carefully slide the bar out and place the bars on the side, out of the traffic area.
Note:
The table should be resting on the floor. You may need to lower or raise the dolly to remove the
dolly ends.
3.) Using the quick release pins, remove each end of the dolly.
4.) Slide the dolly off of the long attachment bar on each side.
5.) Using the quick release pins, remove the two long attachment bars that are attached to the
front and rear table attachment points. Place the bars on the side, out of the traffic area.
6.) Use the 19 mm wrench to remove the bolts on each side of the smaller front table attachment
bracket.Remove the bracket.
7.) Use the 19 mm wrench to remove the bolts on each side of the larger front table attachment
bracket and transporter base. Remove the bracket.
8.) Reassemble the dolly for transportation.
6
Preliminary
Reqs
Figure 1-62
for the location of items in the table dolly.
Chapter 1 - Position Subsystems
O
CT680 S
O
PTIMA
ERIES AND
PTIMA
Procedure
20 min. labor on-site
CT670 I
M
NSTALLATION
ANUAL
Finalization
Page 81

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Optima ct670Optima ct680 quantumOptima ct680 expertOptima ct680 professional

Table of Contents