2 Recommended Power Distribution System

NOTE:
According to the country, the PDB/MDP can be ordered from GEHC.
One configuration 480 V, 60 Hz is certified to UL 508 and CSA C22.2 N°14: North
America. Second configuration (380/400/415 V, 50/60 Hz) is certified to EN 60439-1
and is CE marked (not certified for Japan and China). Please check your country local
regulations for the applicable configuration or refer to your local GE representative.

2.1 Main functions

Lock–out/Tag–out by S1 (Europe) or CB1 handle (US),
Main disconnect device by DMD1 (Europe) or CB1 handle (US),
Low voltage created by TR1 applied on remote, L1 and L2,
Separated injection circuit by TR2/CPT2 and BR,
L is mandatory for French standards,
Protect the Room Emergency Off (AU1, AU2) from accidental activation
168
ANY SAFETY LABELS AFFIXED ON THE PDB AND ON THE EMI FILTER
ENCLOSURE WOULD HAVE TO FULFILL THE LEGAL REQUIREMENT OF THE
COUNTRY CONCERNING THE LANGUAGE TO BE USED FOR INFORMATION
FOR SAFETY ON PROFESSIONAL EQUIPMENT.
To prevent risk of cable overheating or fire.
Local regulations concerning cable tightening inside the PDB must be respected.
The drawing hereafter is a guide to illustrate all functions included inside the main
disconnect room device.
It should be compliant with the specific country rules.
MANDATORY FOR FLUORO UPS OPTION THE PDB FRONT PANEL MUST
WARN THE OPERATOR WITH AN ELECTRICAL HAZARD LABEL.