GE C
OMPANY
D
5472001-1EN, R
IRECTION
EVISION
8.1.8
Warning Light & Door Interlock Connections
8.1.8.1
Warning Light Configuration & Connection
1.) Warning Light is controlled by signals from the system.
2.) This step is site specific. The PDU by default is configured for "no" external warning light
PDU
Figure 2-35 Typical TS6 Warning Light & Door Interlock Connections
Page 130
6
connection. If you have external warning lights, see
EXP_INTLK
PGND
It is recommended that you use the four (4) wire method of adding an X-ray warning light to a
room, as shown in
Figure
-
Minimize EMC interference.
-
Increase contact life of the relay used in the PDU.
Section 8.0 - PDU Cable Connections & Configuration
O
CT680 S
PTIMA
9
10
2-36. When using this method, you:
O
CT670 I
ERIES AND
PTIMA
NSTALLATION
Figure 2-35
for proper connection.
Door Switch
ROOM LIGHT
M
ANUAL