Scan Room Warning Light & Door Interlock
Figure 9-9 Typical TS6 Warning Light & Door Interlock Connections
1
Figure 9-10 TS6 Room Door Interlock Connections - With a Door Interlock
Figure 9-11 TS6 Room Door Interlock Connections - Without a Door Interlock
14
EXP_INTLK
PGND
2
6
3
4
5
7
8
9
Chapter 9 - Interconnection Data
B
S
RIGHT
PEED
9
10
10
If not using a door switch, add a
jumper.
N.O.
Door Switch
I f jumper is not in place ,
exposures will not be made .
C heck this jumper if you get
scan interlock errors .
E
, E
, E
: P
-I
LITE
DGE
XCEL
RE
NSTALLATION
Door Switch