J423 - Patient Ecg Connection; Power Control Pcba; J315 - Power Input From Dc Power Supply; J325 - Smart Battery Connector - Data And Power - GE Responder 2000 Service Manual

Defibrilator/monitor
Hide thumbs Also See for Responder 2000:
Table of Contents

Advertisement

6
/ECG_EN
7
+12V
8
+12V
J423 – Patient ECG Connection
Pin
Number
J4231
RA
J4232
LA
J4233
LL
J4234
V1
J4235
RL
J4236
SHLD

POWER CONTROL PCBA

Responsible for battery charging and DC power switching and distribution. Also includes UI functions
J315 – Power input from DC power supply
Pin
Number
1
GND
2
GND
3
GND
4
+12V_MAIN
5
+12V_MAIN
J325 – Smart Battery Connector – Data and Power
2025653-048 Revision B
ECG Isolated power enable
System +12V (nominal)
System +12V (nominal)
Keystone 4903
Name
Electrode - Right Arm
Electrode - Left Arm
Electrode - Left Leg
Electrode - V1
Electrode - Right Leg
Cable Shield
Molex 39-30-3058
Name
System Ground
System Ground
System Ground
Power Input
Power Input
Suyin 250137MR010G101ZU