Transducer Connection - GE LOGIQ C Series Basic Service Manual

Hide thumbs Also See for LOGIQ C Series:
Table of Contents

Advertisement

GE
D
5273008-100, R
IRECTION
3-5-2-1
Back-end Processor Power Down (cont'd)
2.) The System-Exit window is displayed.
3.) Using the Trackball or Select key, select Shutdown.
4.) The shutdown process takes a few seconds and is complete when the power status LED is turned
amber.
5.) Disconnect the probes.Clean or disinfect all probes as necessary. Store them in their shipping
cases to avoid damage.
3-5-2-2
Scanner Shutdown
Disconnect the Mains Power Cable if necessary. For example: Relocating the scanner.
3-5-3

Transducer Connection

1.) Plug the probe connector into the probe port, then lock the probe.
NOTE:
Please ensure that the probe latch is in an unlocked position before you connect the probe to the
system.
NOTE:
It is not necessary to turn OFF power to connect or disconnect a probe.
9
EVISION
Figure 3-5 System Exit Window
Chapter 3 System Setup
LOGIQ C B
S
S
ASIC
ERIES
M
ERVICE
ANUAL
3 - 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LOGIQ C Series and is the answer not in the manual?

Questions and answers

Addo Godwin Boafo
January 22, 2025

How to divid it into four quadrants

Ramya
April 20, 2025

What is dimmer in ge logic c2 colsole usg machine

User image 680508209ebec
Balaji
April 20, 2025

Where is the dimmer in the logiq USG machine

1 comments:
Mr. Anderson
May 10, 2025

The dimmer (light adjustment) is located on the left side button of the monitor on the GE LOGIQ C Series ultrasound machine.

This answer is automatically generated

Table of Contents