Figure 147 Opening the Front Cover

4. Note the red alignment dots on the handpiece focus connector. Pull back on the metal
connector sleeve (being careful not to pull on the cable) to remove the focus cable
(Figure 148 on page
5. Turn the camera handpiece connector counterclockwise and remove the camera cable
(Figure 148 on page

Figure 148 Removing the Camera and Focus Cables

6. Remove the two screws from the lamp assembly drawer, the smaller panel with the
black handle on the front.
RetCam 3 and RetCam Shuttle User Manual
Pull the front cover out
157).
157).
Camera Handpiece connector
Maintenance and Support
Handpiece Focus connector
PN 18-000393 Rev. A