138
Chapter 10
6. Move the fiber optic cable connector from the white fiber optic (port) connector to the
FA fiber optic (port) connector. (See

Figure 136 FA Front Panel

7. Select Start Timer. During dye injection, start capturing images.
8. Adjust FA light intensity by using the FA light intensity knob (shown in
page
9. You can take still photos or video during the FA session, and you can save still photos
during video replay.
10.When the FA session is completed:
a. Move the fiber optic cable back to the white fiber optic (lamp) connector and switch
b. Remove and store the barrier filter in the special clear plastic box provided, or in a
RetCam 3 and RetCam Shuttle User Manual
Power Switch
NOTE
The timer can be paused or stopped at any time.
138).
NOTE
You cannot adjust the FA lamp light intensity by using the footswitch or
control panel controls.
OFF the FA power.
manner that prevents damage. Clean the filter with compressed air and optical
cleaner and tissue as required.
Fluorescein Angiography (RetCam 3 System)
Figure 136 on page
138.)
White fiber optic (port)
connector
FA Fiber optic
(port) connector
Adjust FA light
intensity
Figure 136 on
PN 18-000393 Rev. A