Alternate Cleaning and Disinfection

• Cleaning Solution (Double): FrankLab Enzymex L9 (immersion time 5 to 15 minutes;
• Disinfection: FrankLab Peraxy-Lens or Peralex H8+ (30 minutes minimum; no longer
Please refer to the specific FrankLab instruction on double cleaning and disinfection.

Figure 141 Inverted Lens Piece Showing Permissible Immersion Depth

Cleaning the Rest of the System

As with any medical device, use good public health practices when handling the
equipment, based on CDC guidelines. In addition, as with typical office equipment, a
gentle wiping with a cloth moistened with mild soap and/or water is recommended. Do not
spray cleaning solutions on any part of the system.

Disposal of Materials

Dispose of waste materials according to local and national requirements. Contact Clarity
Technical Services if additional assistance is required.
RetCam 3 and RetCam Shuttle User Manual
If you clean or disinfect, rinse the lens with sterile water to avoid corneal
de-epithelialization that can be caused by residual solution.
Since disinfection solutions can damage the lens piece, do not exceed the
recommended soak times.
no longer than 2 hours)
than 2 hours)
Maintenance and Support
4mm (0.16")
PN 18-000393 Rev. A
153