Retcam Shuttle Maintenance; Retcam Shuttle Fuse Replacement; Figure 152 Power Entry Module; Figure 153 Opening The Fuse Holder Cover - Clarity RetCam3 User Manual

Ophthalmic imaging systems
Table of Contents

Advertisement

160
Chapter 12

RetCam Shuttle Maintenance

This section describes the specific recommended process for fuse replacement and
illumination lamp replacement for the RetCam Shuttle system. These activities are
appropriate under the following conditions:
• Blown fuse: Replace system fuses.
• Burned out bulb: Replace the illumination bulb.

RetCam Shuttle Fuse Replacement

Figure 152 Power Entry Module

To replace the RetCam Shuttle fuse:
1. Using a flat blade screwdriver, gently pry open the cover of the fuse holder at the top of
the power entry module (see

Figure 153 Opening the Fuse Holder Cover

2. Remove the fuse holder (see
RetCam 3 and RetCam Shuttle User Manual
Do not replace the fuse until the system is completely shut down. Locate the
power entry module on the lower right side of the rear of the cart
on page
160). Make sure to remove the power cord before proceeding any
further.
Figure 153 on page
160).
Figure 154 on page
161).
Maintenance and Support
(Figure 152
PN 18-000393 Rev. A