Figure 3 Computer Electro-Optical (Ceo) Box - Clarity RetCam3 User Manual

Ophthalmic imaging systems
Table of Contents

Advertisement

22
Chapter 1

Figure 3 Computer Electro-Optical (CEO) Box

The Handpiece Focus connector is a direct insert receptacle for insertion of the (white)
focus motor cable, which is part of the handpiece interconnect harness.
To insert the cable, align the red dot on the connector with the red dot on the Focus port
and insert. To detach, pull back on the connector, being careful not to pull on the cable.
The Camera Handpiece connector is a receptacle for insertion of the black camera
controller cable, which is part of the handpiece interconnect harness.
RetCam 3 and RetCam Shuttle User Manual
NOTE
It is not necessary to remove any of the connectors for storage or
transportation.
Do not remove any connections while the system is powered ON. The main
power switch must be OFF prior to connecting or disconnecting any cables
from the CEO box.
Do not look directly into the Lamp (Fiber Optic) opening when the CEO box is
ON.
NOTE
To access the Handpiece Focus connector or the Camera Handpiece
connector, remove the front cover panel.
NOTE
The Handpiece focus connector is not threaded.
Introduction
Fiber Optic Illumination Port
Handpiece focus
Camera Handpiece connector
PN 18-000393 Rev. A