156
Chapter 12
Figure 145 Fuses Removed
.
Replacement fuses are available from Clarity Technical Services.
Figure 146 Fuse Holder Re-Inserted
.
RetCam 3 Illumination Lamp Replacement
Replacement illumination lamps are available from Clarity.
Follow these steps to replace the illumination lamp:
1. If it is not already turned OFF, turn OFF the instrument, according to the instructions
found in
2. Disconnect the main power cord from the wall.
3. Unplug the fiber optic cable from the front panel. Pull out the plastic front cover
(Figure 147 on page
RetCam 3 and RetCam Shuttle User Manual
For continued protection against risk of fire, replace only with same type and
rating fuse.
Appropriately power down and unplug the unit and allow the illumination
lamp to cool before replacing it.
theTurning OFF the RetCam 3 System
157).
Maintenance and Support
section on page 96.
PN 18-000393 Rev. A