To replace the RetCam Shuttle illumination lamp:
1. If not already OFF, turn OFF the Main Power (see
System on page
2. Disconnect the power cord from the wall.
3. Unplug the fiber optic cable from the front panel.
4. Remove the two screws from the lamp assembly drawer.

Figure 156 Front Panel Showing Lamp Assembly Drawer

5. Grip the handle on the lamp assembly drawer. Pull the drawer out slowly. Stop when
the bulb is fully exposed.

Figure 157 Open Lamp Assembly Drawer

6. Just behind the socket and extending from the rear of the illumination lamp is a thick
grey lever, the "lamp ejector" (shown in
RetCam 3 and RetCam Shuttle User Manual
98).
Figure 158 on page
Maintenance and Support
Turning OFF the RetCam Shuttle
Remove 2 screws
Lamp assembly drawer
Lamp assembly drawer handle
163). Be prepared to grab
PN 18-000393 Rev. A